Placing one pacemaker wire can take hours of poking around inside a beating heart, so a Boston University team built a soft robot arm to help.

Today we’re looking at a future where a soft robot with an inflatable arm holds steady inside a beating heart and helps doctors with a fiddly step in fitting a pacemaker. Welcome to FutureProof.

Why this wire is hard to place

In some people with heart failure, the two sides of the heart stop pumping in step with each other. A special type of pacemaker can help. It uses an extra wire to send small electrical pulses that get both sides beating together again.

The hard part is placing the extra wire. It has to go into a vein called the coronary sinus, which runs along the outside of the heart, and doctors reach it from the inside without opening the chest. They slide thin tubes through a vein in the neck or below the collarbone, down into the heart’s upper right chamber, and then look for the small opening that leads into the coronary sinus. The heart keeps beating the whole time, and the only view they get is a live X-ray, where the tools show up as grey shadows.

That opening sits in a different spot from one person to the next, and in some people a small flap of tissue partly covers it. The tubes have a fixed bend at the tip, and the doctor can only push, pull, and twist them from outside the body, which means the tip won’t stay where it’s aimed. So the doctor pokes around with a thin guide wire, again and again, until it slips in.

That search can take anywhere from about 25 minutes to several hours. The researchers point to studies where up to a quarter of patients had trouble at this step, and where it failed outright in up to about 5 in every 100 patients. All that time under X-ray also means radiation for the patient and for the staff in the room.

You might ask why nobody has handed this job to a robot. Robots are already used for other work in the heart and blood vessels, but according to the authors, none has been built for this task. So a team from Boston University, working with MIT and two Boston hospitals, set out to build one. It had to work inside a beating heart, and it had to fit in with the tools doctors already use.

A soft arm with an anchor

So what did they build? The robot is a long, flexible tube that goes in through a vein in the neck. On the way in, it squeezes through a sleeve about 8 millimeters wide, roughly the width of a pencil.

There are two working parts at the end. The first is an anchor, a small basket made of nitinol, a bendy metal that springs back into shape. Once it reaches the large vein just above the heart, the doctor opens it to about three times its closed width, so it braces against the vein wall. Blood keeps flowing through the gaps in the basket.

The second part sits just past the anchor, inside the heart’s upper right chamber. It’s a soft plastic arm with three long, balloon-like chambers inside. Filling the chambers with liquid by different amounts makes the arm bend in any direction and stretch out just enough to reach the target, a few centimeters away.

The arm is hollow all the way down the middle. The doctor pushes their usual wires and tubes straight through it, and the robot points them where they need to go. The operator steers the arm with a handheld controller that drives three syringes on small motors.

The anchor is what lets the arm hold still. A normal tube pivots from where it enters the vein, well away from the heart, so its tip bends and twists out of position. With the anchor braced just above the heart, the pivot point moves close to where the arm is working. The arm can hold its tip in one spot while the doctor probes with the wire, then shift about a millimeter at a time if the first spot is wrong.

What happened in five pigs

The team first tested the robot in a lab model of a heart, which we’ll come back to. They then tried it in five live pigs, whose hearts are close to human hearts in size. Across those animals, they made several attempts to get a wire into the coronary sinus, some with the robot and some with standard tools. They counted how many times the wire poked before it went in, how long each attempt took, and how long the X-ray stayed on.

The most realistic test is the first attempt in an animal, because nobody yet knows where the opening is. That’s the situation a doctor faces with a real patient. With standard tools, the first attempt took about an hour. The wire poked at least 100 times, and the X-ray was on for over 30 minutes.

With the robot, the best first attempt needed just 4 pokes and took a little over 8 minutes, with only a few minutes of X-ray. For that run, the team filled the arm with diluted X-ray dye instead of salt water, so the arm itself showed up on the screen.

There was only one first attempt with standard tools, though, against three with the robot, and the paper doesn’t run any statistical tests on them. The other two robot runs, with salt water in the arm, were slower than the best one but still well ahead of the standard tools.

Once the team had found the opening, they tried again in the same animal, now roughly knowing where to aim. This time, the robot and the standard tools took about the same time and used about the same amount of X-ray.

The poke count stayed very different. Standard tools usually needed 20 to 32 pokes per attempt, while the robot needed 3 or fewer on every one of those later attempts. During the tests, stray touches of the wire against the inside of the heart set off extra, out-of-rhythm heartbeats, and the authors note that extra pokes can risk injuring the heart.

So the slow part is finding the opening the first time, and that’s where the robot saves the most time. The authors put the robot’s advantage down to its stability and fine control. Because it holds still, the wire usually goes in after a few careful pokes instead of dozens.

The robot also delivered a working wire. In one trial, the team passed a pacing wire through it into the vein and switched it on. The pig’s heart rate jumped up to the faster rate they had set, which showed the wire was in place and working.

What this can and can’t tell us

This is an early study, and the paper describes its results as trends. It covers five pigs and a small number of attempts, with no statistical tests. Each pig was only studied during the procedure itself, so the study can’t tell us anything about healing or longer-term effects.

The comparison also wasn’t perfectly even. The robot was driven by a surgical resident, a surgeon in training who had practiced with it, while an expert heart rhythm specialist called out directions from the X-ray screen. That same specialist did the standard procedures.

The device is still on the large side. It needs that 8-millimeter sleeve, and the team wants to shrink it to 6 millimeters. Anything placed in the bloodstream also brings a risk of blood clots, which the authors say future versions may try to reduce. Five of the authors are named inventors on a patent Boston University has filed covering a catheter steering device. In a university news story, the lead researcher estimated it will be at least five to six years before the robot is used in a person.

For now, every prototype is thrown away after one use. One idea the authors are exploring is to make only the soft arm disposable, since it costs just dollars to make, and to clean and reuse everything else. The authors see that as a way to bring this kind of help to hospitals with fewer resources as well as large ones. They also hope a smaller version could reach the heart from other entry points, and one day help with heart surgeries that currently need a heart-lung bypass machine.

Here are the bits we found interesting

  1. The robot is steered with a Nintendo Wii game controller. The team tuned the motors so a quick tap moves the arm’s tip a little over a millimeter, while holding a button sweeps it about a centimeter and a half in under two seconds.

  2. Before the pig tests, the robot was tried inside a real pig heart that had been preserved and fitted with soft robotic muscles that made it beat 60 times a minute, pumping a see-through liquid that flows like blood. A camera inside let the team watch, and the heart’s pressures stayed normal with the robot in place.

  3. In that lab heart, four people who had never used the robot got two minutes of practice before each tried to find the target three times. They and two experienced users all found it in under three and a half minutes.

That’s the paper. If you want to dig into the data yourself, the full study is linked below. But before you go, here’s one stray fact.

The first fully implanted pacemaker went into a Swedish engineer named Arne Larsson on 8 October 1958. Its maker, Rune Elmqvist, used a shoe polish tin as a mold while building it. Larsson went on to use 26 pacemakers over his life and lived to 86.

That’s it for today. We’ll see you in the next one. Cheers.

Source: Rogatinsky, J., Recco, D., Singh, M., Zamora Yañez, L., Matthews, G., Ayers, B., O’Leary, E., Roche, E. T., Hoganson, D., & Ranzani, T. (2025). A beating-heart procedure with soft robotic guidance. Advanced Robotics Research, 1, e2400023.

Source: View Original Paper